I18N Interfaccia tradotta automaticamente. I dati sui ricambi, le specifiche tecniche e le prove documentali restano in inglese; la versione inglese è quella autorevole.

SEZ 02 // SCHEDA RICAMBIO

277222 Miller Flow Meter (Face Shield PAPR)

Verified against primary source · 2026-07-06 other · current · US

Compatibile con questi sistemi

Systems this part fits, with compatibility status and evidence
SistemaStato di compatibilitàProve
Miller Face Shield PAPR T2 Manufacturer-listed part
  1. [01] www.millerwelds.com manufacturer manual Item 14 of Figure 10-2 "Faceshield PAPR Blower Assembly" parts table, Section 10, p.31 of Miller OM-288359A (2021-07): "277222 — Flow Meter, Qty 1". Use procedure in Section 6-8 Testing Airflow, p.20. Archived at sources/miller-o288359a_mil.pdf. 2026-07-06

Specifiche tecniche

Technical specifications
FunctionDisconnect breathing tube from hood, insert flowmeter into breathing tube, hold straight up, and start blower; air flow is adequate if the flowmeter ball moves above the MIN mark (Section 6-8)
Test intervalAlways test air flow before using the respirator; also check monthly if not used regularly (Section 6-8)

STOCKED BY — SUPPLIER CROSS-REFERENCE

Stocked by

See the full supplier directory for all regions and evidence notes.

Listing ≠ endorsement. Confirm availability, approval status and source evidence before purchase or regulated workplace use.

Fonti della scheda

Evidence — dated sources
  1. [01] www.millerwelds.com manufacturer manual OM-288359A Section 6-8 (p.20) and Section 10 (p.31). 2026-07-06